PhoCheck Tiger Technical Specification
Minimum Resolution*
1 ppb or 0.001 mg/m3
Maximum Reading**
20,000 ppm or 20,000 mg/m3
Accuracy*
± 5% display reading ± one digit
Linearity*
± 5% to 100 ppm
Intrinsically Safe Approvals
ATEX II 2G Ex ia IIC T4 (- 15 0C ≤Ta ≤+60 0C), IECEx and CUL
Battery Life
Li-ion: life up to 30 hours, charge time 6.5 hours
Alkaline: 3 x AA, typically 15 hours life
Lamps
10.6 eV Krypton PID lamp (standard) 9.8eV and 11.7eV lamps available
Data Logging*
> 120,000 data log points including date and time stamp (33 hours at 1 log per second)
Communication
USB 1.1
Calibration
Via calibration kit accessory
Alarm
Flashing LED and 95 dBA at 300mm (12”) audible sounder
Selectable vibrating alarm
Pre-programmed TWA and STEL*
Pre-programmed with 450 gases*
Flow Rate
220 ml/min with flow fail alarm
Temperature
Operating: -20 to 60 0C, -4 to 140 0F (non Intrinsically Safe)
Humidity: 0-99% RH (non condensing)
Protection
Designed to IP65 (heavy rain)
C E 1180
EMC tested to EN61326-1:2006, EN50270:2006 & CFR 47:2008 Class A
Weight & Dimensions
Instrument without probe: Width: 340 x Height: 90 x Depth: 60 mm (13.4 x 3.6 x 2.4”)
Standard case: 420 x 320 x 97 mm (16.5 x 12.5 x 3.8”)
Instrument weight 0.72 kg (25 oz)
*Model and gas dependent.